en
es

Deutschland

Italia

España

United Kingdom

United States

Lietuva

Trionyx
Manual
Trionyx

Table of contents

Microbolometer Calibration

 

Available in thermal imaging and Fusion modes

The calibration evens out the microbolometer’s temperature background and eliminates the imperfections of the thermal image (such as vertical stripes, phantom images, etc.) that appear during binoculars operation.

There are three calibration modes: manual (M), semi-automatic (SA) and automatic (А).

Select the desired mode in the menu option Calibration Mode .

Mode М (manual)

  • Close the lens cover (13).
  • Press briefly the ON/OFF (1) button.
  • Open the lens cover (13).

Mode SA (semi-automatic)

  • Press briefly the ON/OFF (1) button to calibrate.
  • You do not need to close the lens cover (the microbolometer is closed with the internal shutter automatically).

Mode А (automatic)

  • The device calibrates by itself according to the firmware algorithm.
  • You do not need to close the lens cover (the microbolometer is closed with the internal shutter automatically).
  • In the automatic mode, the user can calibrate the microbolometer with the ON/OFF (1) (in the SA mode) button.